LIFE (Eclipse): Vaginal Bowel-Control System for Fecal Incontinence

LIFE was a multicenter, open-label, single-arm study of women with at least four minor or major fecal incontinence episodes in two weeks that compared baseline leakage with one month of Eclipse vaginal bowel-control treatment; 48/61 women entering treatment after fitting (78.7%; 95% CI 66–88%) had at least 50% fewer episodes (P<.001).[1]

Design and population
Multicenter prospective open-label single-arm device study. Women aged 19–75 years at six clinical sites with fecal incontinence for at least six months and at least four minor or major episodes on a two-week baseline diary underwent attempted device fitting.
Arms
Eclipse vaginal bowel-control system
Report and follow-up
2015 · One month of treatment, with an optional two-month extension
EndpointResultsEffectFinding
At least 50% fewer episodes at one month among successfully fitted women entering treatment (intention to treat)48/61 (78.7%)95% 66–88; P <0.001Met
At least 50% fewer episodes at three months among women assessed after the optional extension38/44 (86.4%)95% 73–95Descriptive

Harms: Serious device-related adverse events: 0/110; Device-related pelvic cramping or discomfort: 25/110 (22.7%); Device-related adverse events during fitting: 47/110 women (42.7%); 64 mild or moderate events, including urinary incontinence, urgency, or frequency in 11/110 (10.0%) and vaginal findings in 10/110 (9.1%)

NCT01655498 · Report DOI

Design​

Women aged 19–75 years with fecal incontinence for at least six months underwent device fitting at six clinical sites. The primary response definition was at least a 50% reduction in incontinence episodes at one month, tested against a prespecified success rate of 40% using a one-sided α of 0.025. The intention-to-treat analysis started with the 61 women who were fitted and entered treatment; the per-protocol supporting analysis required at least seven days of one-month diary data. An optional two-month extension followed.[1]

Results​

The primary one-month response exceeded the prespecified 40% threshold; the per-protocol response was 48/56 (85.7%; 95% CI 74–94%; P<.001). Among per-protocol women, 23/56 (41.1%; 95% CI 28–55%) had complete continence, and mean episodes per two weeks fell from 11.6 ± 9.5 to 2.1 ± 2.9 (P<.001). At one month, 48/56 (85.7%) rated bowel symptoms “much better” or “very much better.” Among 44 women with analyzable diaries after the optional extension, 38 (86.4%; 95% CI 73–95%) met the episode-response definition at three months.[1]

In a secondary analysis of 56 evaluable women, the mean within-patient proportion of bowel movements accompanied by urgency fell from 54% at baseline to 26% at one month (P<.0001), and the proportion with incomplete evacuation fell from 39% to 26% (P=.0034).[2] Liquid stool (Bristol 6 or 7) fell from 36% to 21% among 55 women with one-month Bristol data (P=.0001); 8/12 women with more than two bowel movements daily at baseline shifted to two or fewer.[2] A fitting analysis associated prior prolapse surgery and shorter vaginal length with unsuccessful fit; prior prolapse surgery was present in 5/61 successfully fitted versus 16/49 unsuccessfully fitted women, and mean vaginal length was 8.6 ± 1.2 versus 8.0 ± 1.1 cm.[3] Women without prior prolapse surgery had 4.7 times the adjusted odds of successful fitting (95% CI 1.53–14.53; P=.007), and the odds increased 1.49-fold per additional centimeter of vaginal length (95% CI 1.02–2.17; P=.041).[3]

Surgery and treatment delivery​

Eclipse consists of a vaginal insert and pressure-regulated pump; users deflated the insert for defecation and were encouraged to remove it weekly for cleaning. Of 110 women who entered fitting, 61 (55.5%) entered treatment and 49 did not: 28 had an unsatisfactory fit, four could not self-manage the insert, one was lost to follow-up, two had logistical or timing constraints, and 14 withdrew.[1] Among 56 per-protocol women, 54 (96.4%) reported that the insert was comfortable or could not be felt.[1] Forty-eight entered the optional extension and 44 had analyzable three-month diaries.[1][2]

The fitting analysis describes 13 round or elliptical base sizes and assessment of comfort, pubic-bone position, retention on Valsalva and rectal occlusion with the balloon inflated. Up to four fitting visits were allowed; 68/110 (61.8%) needed one. Among 49 women who did not complete fitting, 16 failed initial attempts, 29 later reported displacement or discomfort, and four withdrew.[3]

Adverse events​

No serious device-related adverse events were reported. During fitting, 47/110 women (42.7%) had 64 mild or moderate device-related events, including urinary incontinence, urgency, or frequency in 11/110 (10.0%) and vaginal erythema, petechiae, abrasion, irritation, or bruising in 10/110 (9.1%). Device-related pelvic cramping or discomfort occurred in 25/110 women (22.7%); 16/25 (64%) arose during fitting.[1]

Limitations​

The study had no control group. Its treated-population response excludes the 49 women who did not enter treatment after fitting assessment, and the three-month assessment included 44 women after an optional extension.[1]

In conversation​

Same setting NASHA Dx (2011)

Guidelines and approvals​

ASCRS 2023 states that vaginal mechanical inserts are not routinely recommended for fecal incontinence .[4] On February 12, 2015, the FDA granted De Novo classification of the Eclipse System as a class II prescription rectal-control device indicated for fecal incontinence in adult women.[5]

Report history​

LIFE (Eclipse) report history
PublishedReportCutoffHeadline resultsCitation
2015One-month pivotal report · primary—
At least 50% fewer episodes at one month among successfully fitted women entering treatment (intention to treat): 48/61 (78.7%); 95% 66–88; P <0.001 Met
At least 50% fewer episodes at one month among women with a valid treatment diary (per protocol): 48/56 (85.7%); 95% 74–94; P <0.001 Not tested
At least 50% fewer episodes at three months among women assessed after the optional extension: 38/44 (86.4%); 95% 73–95 Descriptive
Bowel symptoms rated very much or much better at one month among women with a valid diary: 48/56 (85.7%); — Not tested
No fecal incontinence episodes at one month in the per-protocol population: 23/56 (41.1%); 95% 28–55 Not tested
Mean episodes per two weeks, baseline to one month in the per-protocol population: 11.6 ± 9.5 to 2.1 ± 2.9; P <0.001 Not tested
Richter HE, Matthews CA, Muir T, et al. "A Vaginal Bowel-Control System for the Treatment of Fecal Incontinence." *Obstet Gynecol.* 2015;125(3):540-547. doi:10.1097/AOG.0000000000000639 DOI
2016Bowel-function analysis · secondary—
Mean within-patient proportion of bowel movements associated with urgency, baseline to one month (per protocol, n=56): 54% to 26%; P <0.0001 Descriptive
Mean within-patient proportion of bowel movements with Bristol stool form 6 or 7, baseline to one month (per protocol; n=56 baseline, n=55 at one month): 36% to 21%; P 0.0001 Descriptive
Mean within-patient proportion of bowel movements with incomplete evacuation, baseline to one month (per protocol, n=56): 39% to 26%; P 0.0034 Descriptive
More than 25% of bowel movements affected at baseline versus one month (per protocol): Liquid stool 34/56 (61%) to 17/55 (31%); urgency 41/56 (73%) to 22/56 (39%); incomplete evacuation 31/56 (55%) to 20/56 (36%); P liquid 0.0002; urgency <0.0001; incomplete evacuation 0.0076 Descriptive
More than two bowel movements per day at baseline, shifting to two or fewer per day at one month: 8/12 (66.7%); — Descriptive
Varma MG, Matthews CA, Muir T, et al. "Impact of a Novel Vaginal Bowel Control System on Bowel Function." *Dis Colon Rectum.* 2016;59(2):127-131. doi:10.1097/DCR.0000000000000517 DOI
2016Fitting-predictor analysis · secondary—
Adjusted odds of successful fit without versus with prior prolapse surgery: Prior prolapse surgery: 5/61 successfully fitted versus 16/49 unsuccessfully fitted; OR 4.7; CI 1.53–14.53; P 0.007 Descriptive
Adjusted odds of successful fit per additional centimeter of vaginal length: Mean vaginal length 8.6 ± 1.2 cm successfully fitted versus 8.0 ± 1.1 cm unsuccessfully fitted; OR 1.49; CI 1.02–2.17; P 0.041 Descriptive
Matthews CA, Varma MG, Takase-Sanchez MM, et al. "Characteristics Associated With Successful Fitting of a Vaginal Bowel Control System for Fecal Incontinence." *Female Pelvic Med Reconstr Surg.* 2016;22(5):359-363. doi:10.1097/SPV.0000000000000290 DOI

References​

1. Richter HE, Matthews CA, Muir T, et al. "A Vaginal Bowel-Control System for the Treatment of Fecal Incontinence." Obstet Gynecol. 2015;125(3):540-547. doi:10.1097/AOG.0000000000000639

2. Varma MG, Matthews CA, Muir T, et al. "Impact of a Novel Vaginal Bowel Control System on Bowel Function." Dis Colon Rectum. 2016;59(2):127-131. doi:10.1097/DCR.0000000000000517

3. Matthews CA, Varma MG, Takase-Sanchez MM, et al. "Characteristics Associated With Successful Fitting of a Vaginal Bowel Control System for Fecal Incontinence." Female Pelvic Med Reconstr Surg. 2016;22(5):359-363. doi:10.1097/SPV.0000000000000290

4. Bordeianou LG, Thorsen AJ, Keller DS, et al. "The American Society of Colon and Rectal Surgeons Clinical Practice Guidelines for the Management of Fecal Incontinence." Dis Colon Rectum. 2023;66(5):647-661. doi:10.1097/DCR.0000000000002776

5. U.S. Food and Drug Administration. Eclipse System De Novo classification order, DEN140020. February 12, 2015.