TITAN 2: Implantable Tibial Neuromodulation for Urgency Urinary Incontinence

TITAN 2 was a prospective, multicenter, open-label, single-arm pivotal study of a leadless rechargeable tibial neurostimulator in 126 implanted adults with urgency urinary incontinence after failure or intolerance of at least two overactive bladder medications; its six-month responder rate exceeded the prespecified performance goal.[1]

Design and population
Prospective, multicenter, open-label, single-arm pivotal device study. 126 implanted adults with urgency urinary incontinence after failure or intolerance of at least two overactive bladder medications at US sites.
Arms
Medtronic leadless rechargeable tibial neurostimulator
Report and follow-up
2026 · Primary outcome at 6 months; safety and observed or imputed efficacy through 12 months
EndpointResultsEffectFinding
At least 50% fewer daily UUI episodes on a 3-day diary at 6 months among complete cases; prespecified performance goal 40%59% (95% CI 49%-67%)P <0.0001Met

Harms: Adverse device effects through 12 months: 25/126 (19.8%); 30 events; Procedure-related implant-site infection: 7%; 1 explant; Serious adverse device effect: 1/126 (0.8%)

NCT05226286 · Report DOI

Design​

The study enrolled 188 people at 26 US sites and implanted 126. Implantation required neither a preliminary stimulation test nor prior percutaneous tibial neuromodulation. Three-day diaries measured daily urgency urinary incontinence (UUI) episodes. A responder had at least 50% fewer episodes at six months. The primary hypothesis tested whether the responder rate exceeded a 40% performance goal, with a one-sided alpha of 0.025. The goal was based on lower confidence limits from sacral neuromodulation studies and had also been used in the eCoin pivotal study.[1]

The six-month efficacy analysis used participants with baseline and follow-up data because missingness stayed below the prespecified 5% imputation threshold. Missingness exceeded that threshold at 12 months, so the later analysis imputed outcomes for the implanted cohort. Safety was assessed in all implanted participants, with related events adjudicated by an external clinical events committee.[1]

Results​

At six months, 59% had at least 50% fewer daily UUI episodes (95% CI 49%–67%; P < 0.0001 versus the 40% goal). At 12 months, the imputed responder rate was 61%. Complete dryness on the diary was reported by 19% at six months and 22% at 12 months.[1]

Among participants with baseline urinary frequency of at least 10 voids per day, mean frequency fell from 12.9 to 10.4 daily voids at six months, a mean reduction of 2.5 (P < 0.0001). The mean Urgency Perception Scale increase was 0.4 points (95% CI 0.3–0.6; P < 0.0001). Mean OAB-q health-related quality-of-life improvement was 27.1 points (95% CI 22.6–31.7) at six months and 25.5 points (95% CI 20.7–30.3) at 12 months.[1]

Surgery and treatment delivery​

Investigators placed the device under local anesthetic in a subcutaneous pocket superficial to the deep fascia over the tibial nerve. Their landmarks were the medial malleolus, posterior tibial border and anterior Achilles border. The article describes a mark 4 cm proximal to the malleolar midpoint, a second mark three-quarters of the way from tibia to Achilles, and an incision and insertion point 5 cm proximal to that mark. A suture loop at the proximal device end permitted retrieval for repositioning; shallow fixation to fascia was optional. The authors state that the commercial implant procedure differs in some respects from this investigational procedure.[1]

All 126 implants elicited a sensory or motor response. Procedures occurred in hospital operating rooms (36/126), ambulatory surgical centers (44/126), and outpatient clinic procedure rooms (46/126). Therapy was to begin within 24 hours, initially programmed for 30 minutes every other day at 20 Hz and a 200-microsecond pulse width, with amplitude set to a comfortable sensory or motor response. Weekly charging was recommended for the investigational device; the article says the commercial device typically requires annual or twice-yearly charging. Follow-up was completed by 125 participants at six months and 118 at 12 months.[1]

Adverse events​

Through 12 months, 25/126 implanted participants (19.8%) had 30 related adverse device effects.[1] The article reported procedure-related implant-site infection in 7%; one wound infection led to device explantation, while the other infections resolved with antibiotics.[1] It identified one serious related event, resolved Clostridioides difficile colitis possibly associated with perioperative antibiotics. The FDA's approval review additionally classified the wound infection requiring explantation as serious.[1][2]

Limitations​

Participants and investigators were unblinded, and the study had no sham or active comparator. The 40% performance goal came from other neuromodulation studies, so its clinical meaning depends on that choice. The published report covers outcomes through 12 months; longer follow-up was planned. Sponsor employees were among the authors, and the trial was Medtronic sponsored.[1]

In conversation​

  • Builds on eCoin pivotal study (2021). TITAN 2 adopted the 40% responder performance goal used in the eCoin pivotal study.

Same setting OrBIT (2009), Finazzi-Agrò PTNS versus sham (2010), SUmiT (2010), OASIS (Revi) (2024), Intibia Pivotal Study (2026)

TITAN 2 used an implanted, automatically scheduled rechargeable stimulator. The Revi system uses an external wearable during stimulation, and eCoin is a nonrechargeable implant; their pivotal studies used different eligibility rules and stimulation schedules.[1]

Guidelines and approvals​

The FDA approved the Medtronic Altaviva system on September 18, 2025, for UUI in patients who failed or could not tolerate more conservative treatments. Its original approval rested on TITAN 2 data through 12 months.[2]

Report history​

TITAN 2 report history
PublishedReportCutoffHeadline resultsCitation
2026-01-26Six- and 12-month pivotal report · primary—
At least 50% fewer daily UUI episodes on a 3-day diary at 6 months among complete cases; prespecified performance goal 40%: 59% (95% CI 49%-67%); P <0.0001 Met
At least 50% fewer daily UUI episodes at 12 months; all implanted with imputation: 61%; —
No UUI episodes on the follow-up diary at 6 and 12 months: 19% at 6 months; 22% at 12 months; —
Mean daily voids among subjects with baseline frequency of at least 10 voids/day, baseline to 6 months: 12.9 to 10.4 voids/day; mean reduction 2.5 voids/day; P <0.0001
Mean increase from baseline at 6 months: 0.4 points (95% CI 0.3-0.6); P <0.0001
Mean improvement from baseline at 6 and 12 months: 27.1 points (95% CI 22.6-31.7) at 6 months; 25.5 points (95% CI 20.7-30.3) at 12 months; P <0.0001
Lee U, Xavier K, Carey J, et al. "Implantable Tibial Neuromodulation Therapy Improves Symptoms of Urge Urinary Incontinence From the TITAN 2 Pivotal Study." *J Urol.* 2026;215(6):702-712. doi:10.1097/JU.0000000000004958 DOI

References​

1. Lee U, Xavier K, Carey J, et al. "Implantable Tibial Neuromodulation Therapy Improves Symptoms of Urge Urinary Incontinence From the TITAN 2 Pivotal Study." J Urol. 2026;215(6):702-712. doi:10.1097/JU.0000000000004958

2. US Food and Drug Administration. Medtronic Altaviva System: Premarket Approval P240011. September 18, 2025. FDA approval record; Summary of Safety and Effectiveness Data.