Skip to main content

Harmonic ACE Ultrasonic Shears (da Vinci)

Harmonic instruments cut and coagulate tissue using ultrasonic blade motion. The 2010-cleared da Vinci Harmonic ACE assembly and the Xi/X catalog's single-use 8 mm Harmonic ACE curved shears (480275) are different robotic configurations; handheld ACE+7 is a separate product.[1][15][2] Their platform compatibility, vessel limits and disposable components must be checked separately.[1][2] Intuitive's January 2026 multiport catalog lists the 480275 shears and states that Harmonic ACE is not available on da Vinci 5.[16]

Identify the Configuration​

ConfigurationWhat the primary device documents establish
Historical da Vinci Harmonic ACEFDA clearance K093217 describes 5 and 8 mm configurations for specified earlier da Vinci systems, with a reusable proximal housing and disposable distal shaft/grip insert, connected to the specified Ethicon generator and handpiece.[1]
Handheld/laparoscopic ACE+7A separate instrument with a 7 mm vessel-coagulation indication in Advanced Hemostasis mode. This indication must not be transferred to a legacy robotic ACE or to every Harmonic model/mode.[2]

The traditional robotic Harmonic lacks the wristed tip of many da Vinci graspers and scissors. Bedside use of a compatible handheld instrument through an assistant port is the usual workaround. It keeps the robotic arms free for wristed instruments. In a 10-case robotic gastrectomy series, console time was 251 minutes with assistant-port ultrasonic shears and 306 minutes with robotic devices only (single centre, not randomized). Check the actual system, instrument availability and team roles.[1][3][14]

Mechanism and Thermal Effects​

Ultrasonic motion plus tissue compression produces cutting and coagulation without using the patient as an electrosurgical RF current circuit. A patient return electrode is not required for this ultrasonic action.[2][4] The blade and passive jaw nevertheless become hot.[2][4]

In an experimental study using bovine mesentery and lamb veins, the tested ACE blade reached about 191°C during cutting and took 35.7 seconds to cool to the study's 60°C endpoint; the passive jaw took 25.4 seconds.[4] Another tested device had a longer active-blade cooling time, so ACE was not universally the “slowest to cool.”[4]

Avoid contact between recently activated surfaces and vulnerable adjacent structures. No single published millimeter margin establishes safety beside a nerve, ureter or bowel.

Vessel Sealing: Bench Data​

Vessel size affects seal performance, but the labeled vessel limit for the exact instrument and activation mode remains the operative ceiling.[2] These studies used harvested porcine arteries, in vivo porcine vessel sealing followed by ex vivo burst testing, or ex vivo human colorectal specimens; none estimates clinical failure rates.[8][9][10]

StudySettingFinding
Harold 2003[8]Porcine arteries, 16 per device per size groupMean burst pressure, electrothermal bipolar sealer versus ultrasonic shears: 601 versus 205 mmHg at 4–5 mm and 442 versus 175 mmHg at 6–7 mm. For the bipolar sealer, both size groups had higher burst pressures than the 2–3 mm group (128 mmHg; p = 0.0001). Burst pressures for the two devices at 2–3 mm did not differ significantly.
Okhunov 2018[9]246 porcine vesselsHarmonic Scalpel Ace Plus had the lowest mean burst pressure among five devices: 571 mmHg in medium (5.1–7 mm) and 254 mmHg in large (7.1–9 mm) arteries. Seal failure occurred in 20% of medium and 40% of large arteries. Harmonic Ace +7, Caiman 5 and LigaSure had no failures up to 9 mm. Failure was defined as a burst pressure below 300 mmHg for arteries; reported group mean burst pressures were all above 250 mmHg.
Noble 2011[10]93 ex vivo mesenteric vessels, mean diameter about 1 mmHarmonic Ace, LigaSure and LOTUS burst pressures did not differ significantly. Depth of thermal damage was 1.95 mm with Harmonic Ace and 3.37 mm with LigaSure (p < 0.001).

Use ultrasonic shears alone for small vessels and mesentery. For arteries above about 5 mm, use a sealer rated for that size, clips or ligatures. These data used older handheld models, so check the current instructions for the exact device.

Reconstructive Applications​

Potential tasks include selected soft-tissue and small-vessel division during abdominal exposure or diversion. Preserve bowel and ureteral vascular supply, isolate the intended tissue and use the vessel limit for the actual model and activation mode. Clips, ligatures or other vascular control remain appropriate when the tissue or vessel is outside that indication.

Cold sharp dissection and selective nonthermal hemostasis may be preferable around neurovascular bundles, ureter, fine reconstructive planes or prosthetic components. Neither ultrasonic nor bipolar energy has a universal safety advantage for these structures.

Technique and Device Instructions​

The ACE+7 instructions illustrate why the model and mode matter: power, tissue tension, grip, blade features and tissue type all affect the result; its 7 mm indication specifies Advanced Hemostasis. The instructions prohibit contact with metal while activated and abrasive cleaning of the blade.[2] Use the exact device instructions for assembly, cleaning, cooling and activation.[2]

Keep the working end visible, identify nearby structures and avoid activating on unintended tissue.[5] Ultrasonic devices can generate smoke or vapor; a small randomized colpotomy study found better visibility than monopolar energy.[5]

Evidence and Choice of Energy Device​

Kong's 2017 robotic gastrectomy feasibility study compared 17 cases with the articulating EndoWrist Vessel Sealer against 52 concurrent cases with ultrasonic shears. It was prospective but not randomized. Operative time, blood loss and complications did not differ. C-reactive protein on day 2 (8.06 versus 11.7) and albumin on day 5 (3.51 versus 3.32) favored the sealer.[6] In a randomized three-arm trial of laparoscopic distal gastrectomy (n = 174), day-2 CRP was lower with advanced bipolar sealing than with ultrasonic shears (9.03 versus 11.12), as was blood loss (26.3 versus 43.7 mL); operative time, hospital stay and complications did not differ.[12] These are gastric-surgery surrogates.

Randomized colorectal data favor either energy device over monopolar scissors. In 61 patients having laparoscopic left colectomy, dissection took a median 90 minutes with ultrasonic shears, 105 with a bipolar sealer and 137 with monopolar scissors (p < 0.001).[11] The authors estimated lower cost with either device at 200 cases a year.[11] An umbrella review of ten systematic reviews in surgical oncology reported 25–29 minutes shorter operative time with Harmonic devices, with very low to moderate certainty. Its authors were employed or sponsored by the manufacturer.[13]

The older Cochrane review of laparoscopic colectomy included six heterogeneous trials with 446 participants and could not determine an overall best energy instrument. Device and operating-room costs depend on local use and procedure.[7]

For current robotic sealers, remember the separate limits: SynchroSeal up to 5 mm; Vessel Sealer Extend up to 7 mm. A product's vessel limit and articulation are selection factors; see the vessel-sealer comparison.

See also: Energy Devices, LigaSure, Maryland Bipolar, Monopolar Curved Scissors.

References​

1. US FDA. Da Vinci Harmonic ACE Device, K093217. 2010. 510(k) summary and indications.

2. Ethicon / J&J MedTech. Harmonic ACE+7: Advanced Hemostasis indication and instructions excerpt. Manufacturer product information; brochure with instructions excerpt.

3. Hirahara N, Matsubara T, Hayashi H, Tajima Y. Features and applications of energy devices for prone robot-assisted minimally invasive esophagectomy: a narrative review. J Thorac Dis. 2022;14(9):3606–3612. doi:10.21037/jtd-22-559.

4. Kim FJ, Sehrt D, da Silva RD, et al. Evaluation of emissivity and temperature profile of laparoscopic ultrasonic devices (blades and passive jaws). Surg Endosc. 2015;29(5):1179–1184. doi:10.1007/s00464-014-3787-0.

5. Choi C, Do IG, Song T. Ultrasonic versus monopolar energy-based surgical devices in terms of surgical smoke and lateral thermal damage (ULMOST): a randomized controlled trial. Surg Endosc. 2018;32(11):4415–4421. doi:10.1007/s00464-018-6183-3.

6. Kong SH, Kim TH, Huh YJ, et al. A feasibility study and technical tips for the use of an articulating bipolar vessel sealer in da Vinci robot-assisted gastrectomy. J Laparoendosc Adv Surg Tech A. 2017;27(11):1172–1179. doi:10.1089/lap.2017.0093.

7. Tou S, Malik AI, Wexner SD, Nelson RL. Energy source instruments for laparoscopic colectomy. Cochrane Database Syst Rev. 2011;(5):CD007886. doi:10.1002/14651858.CD007886.pub2.

8. Harold KL, Pollinger H, Matthews BD, et al. "Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries." Surg Endosc. 2003;17(8):1228–30. doi:10.1007/s00464-002-8833-7

9. Okhunov Z, Yoon R, Lusch A, et al. "Evaluation and comparison of contemporary energy-based surgical vessel sealing devices." J Endourol. 2018;32(4):329–37. doi:10.1089/end.2017.0596

10. Noble EJ, Smart NJ, Challand C, et al. "Experimental comparison of mesenteric vessel sealing and thermal damage between one bipolar and two ultrasonic shears devices." Br J Surg. 2011;98(6):797–800. doi:10.1002/bjs.7433

11. Hubner M, Demartines N, Muller S, et al. "Prospective randomized study of monopolar scissors, bipolar vessel sealer and ultrasonic shears in laparoscopic colorectal surgery." Br J Surg. 2008;95(9):1098–104. doi:10.1002/bjs.6321

12. Park JH, Kong SH, Berlth F, et al. "Comparison of perioperative outcomes between bipolar sealing, ultrasonic shears and a hybrid device during laparoscopic gastrectomy for early gastric cancer: a prospective, multicenter, randomized study." Gastric Cancer. 2023;26(3):438–50. doi:10.1007/s10120-023-01365-6

13. Cheng H, Clymer JW, Sadeghirad B, et al. "Performance of Harmonic devices in surgical oncology: an umbrella review of the evidence." World J Surg Oncol. 2018;16(1):2. doi:10.1186/s12957-017-1298-x

14. Kakeji Y, Kuroda D, Nakamura T, et al. "Ultrasonic shears assistance can shorten the console time in robotic gastrectomy for early gastric cancer." BMC Res Notes. 2015;8:443. doi:10.1186/s13104-015-1432-1

15. Intuitive Surgical. Da Vinci X/Xi System Instrument and Accessory Catalog. December 2023;7. Manufacturer catalog.

16. Intuitive. Da Vinci Multiport Instrument and Accessory Catalog. BUS00143 V3 US, January 2026. Manufacturer catalog.